Model Number |
HS-3041EO |
Features
-
The sterilization with gaseous agents serves as the complementary with steam sterilization system used for sterilizing the medical device and materials defying the steam sterilization due to its vulnerability to the high temperature and moisture. Up to now, it is recognized as the sole sterilization system which can replace the steam sterilization method; it is an indispensable equipment for the hospital and the research institutes related with the microorganism.
-
Comtrols
The Controls System with a built-in microprocessor, controls process of the cycle, the displaying of a large 5"LCD showing the operating conditions of the sterilizer, the self-diagnosis which detects abnormality and takes necessary actions and the printer which printing out the cycle performance records automatically.
-
Cycle program
As for the sterilization gas, three(3) kinds of gas mixture with carbon dioxide(CO2) and two(2) Freon (CFC, HCFC) can be used and thus the option for using gas is rather wide; it has the built-in four(4) kinds of basic cycles, one(1) test cycle and user cycle whereby the high temperature (56℃)and the low temperature (38℃)can be selectively used.
The basic cycle is a program generally applied to most to most items to be sterilized. The test cycle is a vacuum leak test for airtightness of the chamber.
The user cycle can be used and saved by changing the temperature, pressure, sterilization time and aeration time at the user's option.
-
Self diangnosis
The Self-Diagnosis detects 14 malfunctions by itself and displays the error codes and gives the warning signal sound.
- Vacuum pump
The vacuum pump is of the diaphragm type with low noise, powerful exhaustion, strong durability and thermal protector.
- Printer
It prints a cycle performance record which reviews, analyzes and saves sterilization results related to management. This document includes sterilizer number, date and year, setting value (sterilization temperature, exposure time, and drying time), performance data of each phase (start and finish time, temperature and pressure), maximum / maximum/minimum temperature and pressure, elapse time of each phases, acceptance or unacceptance of the cycle and other data.(cycle time, cycle number, error codes, power failure/recovery time)
Specifications
| Overall |
W614 xH484 xD792mm |
Cabinet |
W614 xH442 xD646mm |
Chamber
|
W300 xH300 xD450mm, Capacity: 40.5 liter |
Reservoir |
Capacity : 3 Liter |
Operating Temperature |
121°C∼132°C |
Basket |
Rectangular |
Sterilization Agents
|
Mixtures Of Ethylene Oxide and Inert Gases
12% EO + 88% CFC, 10% EO + 90% HCFC,
20% EO + 80% CO2 |
Power source |
AC 220V, 50/60 Hz |
Power Consumption |
1.600W |
Control |
Microprocessor |
Vacuum Pump
|
Flow Rate : 11 liter/min
|
Display
|
Dot Graphic Display LCD
|
| Printer
|
Dot Matrix, 40 Columns, Single Color |
Net Weight |
91kg |
|
|
|
|
| |
|
|
|