Click to enlarge image
|
Model Number |
HRB-60 |
Features
-
This unit forms ideal conditions, which enable long-term, effective storage of biologicals. A refrigerator for such purpose requires uniform distribution of the termperature and precise termperature control to prevent the stored materials from deteriorating or being frozen. This refrigerator is one of the fundamental equipment in blood center, medical institutions, research institutes, and distributors of medicines.
-
It maintains uniform temperature and is free from frost because cooled air coercively circulated.
-
It is outstanding in cold retaining: it maintains the temperature of stored blood at 6°C or below for more than one hour or at 10°C or below for more 2 hours even when power fails at the surrounding temperature of 30°C.
-
The temperature/power monitor gives a digital display of the inner temperature during operations: if the temperature deviates from the setting or the power supply has failed, it gives a digital display of the present temperature, High(Hi),Low(Lo)and power failure(AC) respectively and sounds the warning.
-
The temperature recorder describes for 7days inner temperature in graphs on the recording chart and it must be kept for reference.
Specifications
| Exterior |
W610 x H915 x D635mm, Stainless Steel Plate |
Interior
|
W530 x H528 x D565mm, Stainless Steel Plate
158 liter cap. (5.56cuft), 450mliter pk. x 60 ea |
Drawers
|
2ea, Stainless Steel Plate |
Condensing Unit |
1/5HP, Hermetically Sealed Compressor |
Cooling Unit |
Automatic off-cycle Defrosting system |
Refrigerant
|
R-12 |
Sterilization Agents
|
Mixtures Of Ethylene Oxide and Inert Gases
12% EO + 88% CFC, 10% EO + 90% HCFC,
20% EO + 80% CO2 |
Power source |
AC220V, 50/60㎐ |
Operating Temperature
|
2°C∼4°C |
Monitor Model
|
HTM-6974D |
Net Weight
|
87㎏ |
Optional
|
Temperature Recorder |
|
|
|
|
| |
|
|
|